Print Back
These medical grade isolation boxes are built using a toroidal transformer construction that inherently helps reduce stray fields, increases efficiency and minimizes size compared to traditional EI transformers. Built with Class F (155º) insulation system. The enclosure is made of steel and powder coated white.
Unit of Measure

Check Stock

Click here to check stock >>

Electrical Specifications at 25 Degree Celsius (ºC) Temperature

Maximum Power

1000 V·A
Input Voltage1 240 V
Output Voltage2 120 V

Input Voltage Tolerance (Power Cord)

±10 %

Frequency (Power Cord)

50 Hz60 Hz

Note for Output Voltage

Alternating Current (AC) Output Voltage (Receptacle) at Maximum 8.33 A.

Typical Voltage Regulation from Full Load to No Load

5 %

Alternating Current (AC) Input to Output Voltage (Hi-Pot)

3000 V

Alternating Current (AC) Input and Output to Enclosure Voltage (Ground) (Hi-Pot)

1500 V

Typical Efficiency at Full Load

94 %

Operating Ambient Temperature

Not to Exceed 60 ºC

Leakage Current

Less Than 50 µA

Typical Leakage Current

10 µA

Fusing

Replace only with 6.0 A, 250 VAC, Slow Blow, UL Listed Fuses.

Specifications

Safety

Medical

Length

7.17 in182 mm

Width

9.06 in230 mm

Height

5.75 in146 mm

Weight

21 kg

Warranty

5 years

Enclosure Material

Steel

Enclosure Finish

Powder Coated

Enclosure Color

White

Mounting

Rubber Feet for Horizontal Mounting.

Insulation Class

Class F (155º)

Industry Standards

CSA cULus Listed RoHS 2015/863/EU

Connections

Input Connections

11 ft. Medical Grade Power Cord with NEMA 6-15P Plug.

Output Connections

Duplex Receptacle. NEMA 5-15R.

Note

Note

Medical isolation transformer box is certified to UL 60601-1, 2nd edition medical standard and CSA-C22.2.No. 601.1M90, 2005
UL: (E352717), UL60601-1, 2nd edition and CAN/CSA-C22.2 No. 601.1M90, 2005

Not for use in life critical application.
  • 1 Alternating Current (AC) Input Volatge (Power Cord)
  • 2 Combined loads connected to receptacle (s) must not exceed maximum current rating.