These medical grade isolation boxes are built using a toroidal transformer construction that inherently helps reduce stray fields, increases efficiency and minimizes size compared to traditional EI transformers. Built with Class F (155º) insulation system. The enclosure is made of steel and powder coated white.
Unit of Measure

Check Stock

Click here to check stock >>

N/A

Electrical Specifications at 25 Degree Celsius (ºC) Temperature

Maximum Power

N/A 2400 V·A
Input Voltage1 N/A 240 V
Output Voltage2 N/A 120 V

Input Voltage Tolerance (Power Cord)

N/A ±10 %

Frequency (Power Cord)

N/A 50 Hz60 Hz

Note for Output Voltage

N/A Alternating Current (AC) Output Voltage (Receptacle) at Maximum 16.66 A.

Typical Voltage Regulation from Full Load to No Load

N/A 5 %

Alternating Current (AC) Input to Output Voltage (Hi-Pot)

N/A 3000 V

Alternating Current (AC) Input and Output to Enclosure Voltage (Ground) (Hi-Pot)

N/A 1500 V

Typical Efficiency at Full Load

N/A 94 %

Operating Ambient Temperature

N/A Not to Exceed 60 ºC

Leakage Current

N/A Less Than 50 µA

Typical Leakage Current

N/A 10 µA

Fusing

N/A Replace only with 12.0 A, 250 VAC, Slow Blow, UL Listed Fuses.

Specifications

Safety

N/A Medical

Length

N/A 8.74 in222 mm

Width

N/A 10.63 in270 mm

Height

N/A 6.82 in173 mm

Weight

N/A 42.5 kg

Warranty

N/A 5 years

Enclosure Material

N/A Steel

Enclosure Finish

N/A Powder Coated

Enclosure Color

N/A White

Mounting

N/A Rubber Feet for Horizontal Mounting.

Insulation Class

N/A Class F (155º)

Industry Standards

N/A CSA cULus Listed RoHS 2015/863/EU

Connections

Input Connections

N/A 11 ft. Medical Grade Power Cord with NEMA 6-20P Plug.

Output Connections

N/A Duplex Receptacle. NEMA 5-20R.

Note

Note

N/A Medical isolation transformer box is certified to UL 60601-1, 2nd edition medical standard and CSA-C22.2.No. 601.1M90, 2005
UL: (E352717), UL60601-1, 2nd edition and CAN/CSA-C22.2 No. 601.1M90, 2005

Not for use in life critical application.

SnapMagic

N/A

  • 1 Alternating Current (AC) Input Volatge (Power Cord)
  • 2 Combined loads connected to receptacle (s) must not exceed maximum current rating.